更全的杂志信息网

Upgrade of macromolecular crystallography beamline BL17U1 at SSRF

更新时间:2016-07-05

1 Introduction

Macromolecular crystallography(MX)is the primary technique to determine the atomic structure of biological molecules.The MX beamline plays an important role in this field.With continuing advances in beamline instruments and experimental methods,the data collection on MX beamline has become routine[1].However,a signification portion of samples from challenging projects still diffract poorly.The samples require more specialized experimental conditions,such as a micron-sized beam to match the size of the crystal or of the good diffraction region,a high-precision or specially configured diffractometer,or a noise-free detector to improve the signal-tonoise ratio.To provide these conditions,more attention needs to be paid to the instrument design and integration of the endstation devices.

BL17U1 is the first dedicated MX beamline at Shanghai Synchrotron Radiation Facility(SSRF).In the 8 years since it opened to users in 2009,this beamline has produced more than 2200 macromolecular structures[2].This beamline employs an in-vacuum undulator as the source.The optical setup is designed to be simple and stable to achieve a high- flux focused beam with a reasonable beam size and small divergence[3].The endstation of the beamline receives continuous partial upgrades;for example,an active beamstop was developed[4],and a microcollimator is now used to rapidly select micron beams,but the overall performance of the endstation devices is dif ficult to improve.

The upgrade reported here aims to improve the overall performance of the BL17U1 endstation.The key parameters of the devices will be improved.For example,the sphere of confusion (SOC)of the diffractometer is expected to reach ≤ 1 μm peak to peak.The EIGER X 16M detector(Dectris Ltd.)will allow data collection at up to 133 Hz.To exploit the full speed of the detector,the reproducibility and precision of the sample stages will be speci fied to meet the requirements of the centering and data collection schemes.The reliability of beamline operation,apart from the improvement of the instrument performance,is also considered carefully.

The installation was carried out during the summer shutdown,and final testing was conducted in November 2017.Before the goniometer was installed,it was assembled and tested in the development lab,where the vital components underwent a site acceptance test.On the basis of the of fline measurement result,the assembly was revised several times to obtain the best performance.During the shutdown,the previous goniometer was removed,and the home-integrated one was transported to the site.The EIGER X 16M was delivered to the beamline in October and installed immediately.The upgraded endstation was tested using test samples.The design and realization of this upgraded endstation are presented in this paper.

2 Hardware

The goniometer and detector are supported on separate tables to eliminate their effects on each other.The goniometer includes a single rotatory axis,a video system for observing the sample,and the shutter and beam condition components.The configuration of the hardware is suitable for manual and robotic sample changes.To reduce the effect of thermal drift,the temperature and humidity are further controlled in the endstation.

2.1 Goniometer

Then,the two-stage temperature control is adopted for the high-precision control system.Temperature control is achieved primarily by the equipment itself,and the temperature is controlled within 24.5± 0.2°C when the temperature is set to 25°C.The secondary precision control unit,which was installed at the terminal of the wind pipe,is also used,and finally a temperature of 25.0± 0.1°C is obtained in the EH.The temperature setting is changeable,and a program optimizes the twostage control automatically.

In order to align the center of the on-axis microscope with the center of the X-ray beam,a five-axis motion system with x,y and z movement and pitch and yaw rotation is chosen,as shown in Fig.4.In the endstation of the SSRF-BL17U1 beamline,the actual X-ray beam shape can be observed using an yttrium aluminate garnet plate.A crosshair is drawn to represent the center of the on-axis microscope.Thus,the beam center and on-axis microscope center can be easily aligned by changing x,z,the pitch,and the yaw.The y direction motion is used for adjusting the focus of the on-axis microscope.In order to make the image much clearer,a backlight is positioned after the sample holder by a rotary actuator.During data collection,the backlight is moved out of the X-ray beam.

式中,wk-1、vk分别为n维和m维的高维的高斯白噪声向量序列:wk→WN(qk,Qk),vk→WN(rk,Rk),这里的噪声均值qk和rk不一定为零,并且假设两噪声之间互不相关,即有:

The high-precision and high-speed sample centering system uses the sample_x,sample_y,and sample_z motions.The sample_x and sample_y motions move the sample in the vertical plane,and the sample_z motion moves the sample in the lateral direction.As shown in Fig.1,there is a two-dimensional movement stage on top of the air-bearing rotary stage(along the axial direction),and these two motions are labeled sample_x and sample_y.SLC series piezo-drive positioners from SmarAct are used for the sample_x and sample_y motions because of the high resolution of up to 1 nm or less of the new picoscale interferometer.There is also a two-dimensional stepper stage below the air-bearing rotary stage,the horizontal and vertical motions of which are de fined as sample_z and gonio_vert,respectively.The resolution of these two stages is 0.5 μm.The sample_x,sample_y,and sample_z motions are used to move the sample to the point of the beam center,and the gonio_vert motion is used to move the rotation axis of the rotary stage to the microscope crosshair or beam center.

Fig.1(Color figure online)Design(a)and photograph(b)of the newly built ultrahigh-precision diffractometer in the SSRF-BL17U1 endstation

Fig.2(Color figure online)Setup of the SOC measurement(a)and test results(b)of ABRT 260 air-bearing rotary stage,which show that the SOC is less than ± 0.5 μm

For the customized on-axis microscope from Qioptiq Company,the magni fication of the eyepiece is 10×,and the magni fication of the objective lens is changeable,with a maximum of 12×for a digital camera.The resolving power of the on-axis microscope is measured using a highresolution target,1951 USAF.The microscope can clearly distinguish element 1 in group 9 under high magni fication,as shown in Fig.3,indicating that its resolution could be as low as 1 μm.Thus,this on-axis microscope is highly suitable for crystals smaller than 5 μm.

Fig.3 USAF 1951 target under the on-axis microscope.(Width of line in group 9,element 1 is 0.98 μm)

An ABRT 260 air-bearing direct-drive rotary stage(Aerotech, Inc.) was used for the high-precision goniometer.This stage has superior angular positioning,velocity stability,and error motion performance along with an impressive payload capacity and outstanding radial and axial stiffness.Further,this stage not only has excellent axial and radial error motions,but also has outstanding tilt error motion,which is very important for our goniometer.In the synchronous model,the tilt error motion is< 2.4 μrad,which means that the SOC could be less than± 0.5 μm,as the axial distance is less than 0.2 m.Figure 2 shows the test results of the ABRT 260 air-bearing stage using a DT6220-CS02 capacitive sensor(Micro-Epsilon),which show that the horizontal and vertical SOC are both< ± 0.5 μm.

当时的公狗是咱现在的院长。那时候,他在葫芦套村当村长。他和老鳜鱼的老婆有一腿。老鳜鱼的老婆叫三妮。在葫芦套村是数得着的俊媳妇。

An attempt to hide the hardware details in the graphical user interface was made by developing several hardware services.The SSRF’s revised version of Blu-Ice has been implemented and provided to general users since 2010[6].Users will not see any operational difference after the hardware is upgraded.The integration of the diffractometer and EIGER detector is described below.

Fig.4(Color figure online)Five-axis motion system for aligning onaxis microscope and beam center

Fig.5(Color figure online)Photograph of the upgraded endstation of SSRF-BL17U1;on the left is the newly built high-precision goniometer,and on the right is the EIGER X 16 M detector

2.2 Temperature and humidity control

A dry and thermally stable environment is very important for microbeam MX experiments.There are many instruments inside the BL17U1 endstation hutch(EH),such as the high-precision diffractometer and EIGER X 16M detector,that require temperature and humidity control in the BL17U1 EH with a temperature stability of± 0.1 °C and a low relative humidity of less than 30%.

2.2 组织病理学改变 A组造模后,光镜下6h左右可见粘膜下层或固有层严重分离,肠绒毛腺体崩解分离、脱落缺失;12h部分肠绒毛上皮细胞开始修复,但轮廓不清;B组病理组织学改变较A组轻微,2h黏膜下层和肌层仅轻度水肿;6h粘膜下层或固有层轻中度分离,肠绒毛结构基本存在;12h修复明显。C组和D组肠绒毛完整,两组无明显不同。各组新生鼠肠组织损伤病理评分见表1。

The temperature in the experimental hall at SSRF is already controlled at25± 3°C,butthe humidity requirement has not been considered.In particular,the humidity is very high in the plum rainy season in Shanghai from April to June.In MX experiments,a moist environment will cause icing not only of the crystal samples,but also of some devices such as the robot gripper.To reduce the thermal fluctuation and to obtain a low-humidity environment,a dedicated air-conditioned facility has been designed and installed for the BL17U1 EH.

The hardware of the diffractometer control system consists mainly of an Aerotech multi-axis motion controller(Ensemble),a SmarAct modular control system(MCS),a Kohzu SC series motion controller,a Galil DMC series motion controller,and some independent external functionalmodules.The hardware structure ofthe goniometer control system is shown in Fig.7.Each controller serves the motion control of its own equipment and uniformly outputs useful signals to the logical I/O module of the Galil controller,which is programmable to handle the input signals and outputs accurate control signals to other devices or external functional modules.

As shown in Fig.1,the newly built ultrahigh-precision diffractometer consists mainly of an air-bearing goniometerwith atilterrormotion smallerthan< 2.4 μrad,a high-precision sample centering system with a repeatability of±25 nm,an on-axis microscope with high magni fication(12×10)and a long working distance(~33 mm),a backlight,an ultrafast shutter,a movable beamstop with an intensity monitor,changeable collimators,and other associated equipment such as a cryostream and fluorescence detector.Owing to the advantages of the ultrahigh-precision air-bearing goniometer,high-precision sample centering mechanism,and high-magni fication onaxis microscope,the newly built goniometer can achieve high accuracy and high speed in finding and positioning ultrasmall(< 5 μm)crystals at the beam center.

将产生的感应电势经过隔离开关、电压冲击保护、滤波、隔离变压器等组件,送入电源变换模块。电源变换模块利用整流、电压变换等电路把感应的交流电压(9 V~100 V)转换为系统供电所需的12 V直流电。

The temperature-and humidity-controlled EH has operated well with stable performance for 4 months.The temperature and humidity data obtained for 2 days(Feb.6–7,2018)are shown in Fig.6b.We found that the humidity inside the hutch is clearly affected by the status of the door(whether it is open or closed).However,the temperature is affected only slightly within a short door opening time,that is,less than 15 min.The reason is that the temperature difference between the SSRF hall and the hutch is very small.Therefore,the newly built air-conditioning system has realized temperature and humidity control for the hutch,maintaining a constant temperature(23± 0.1°C)and low humidity(less than 30%RH)successfully.

Vibration during operation of the air-conditioning system was considered,and it can be isolated.First,we used a flexible connection between the outlet of the fan and the air pipe.Second,the internal equipment,which consists of the fan,three coil refrigeration units,and motor,was placed on rubber shock absorbers.Thus,the vibration to the diffractometer through the ground is negligible.

2.3 Area detector

The EIGER X 16M hybrid pixel area detector has an active area 311.2 mm in width and 327.8 mm in height.The small pixel size of 75 μm and the fact that the pointspread function is limited to one pixel also make the detector suitable for large-unit-cell samples.The maximum frame rate is 133 Hz,whereas the rate for the region of interest can be as high as 750 Hz.The frame readout is continuous.The format of diffraction images is HDF5[5],and images can be compressed for data transmission.All of these features are very important for the new data collection schemes.

The detector is installed on the original support table,which can drive the detector in six dimensions.A minimum distance of 120 mm can be obtained under the current setup.At this distance,the recordable resolution is 1.1 A˚at an incident X-ray wavelength of 0.97 A˚.The resolution will be 0.78 at 0.69 A˚(X-ray energy of 18 keV)and 2.81 at 2.48 A˚.The minimum distance could be further reduced after the redesign of the transfer board and the detector’s protective cover.

Fig.6 Schematic diagram of the low-temperature chilled water method(a),temperature and humidity data obtained for 2 days(Feb.6–7,2018)(b),and distribution of the temperature data(c)

3 Control system

In order to simultaneously block the transmitted X-rays and monitor the intensity of the X-rays after the sample,a beamstop with an intensity monitor was designed and manufactured and is shown in Fig.1.A changeable collimator with pinholes of different sizes(10,20,50,100,and 200 μm)is installed before the sample to provide beams of different sizes for diffraction experiments.In addition,a fast shutter is mounted before the on-axis microscope along the X-ray beam to control the exposure time in diffraction experiments.Figure 5 shows the installation of the newly built high-precision goniometer with the new EIGER X 16M detector.

The low-temperature chilled water method adopted for the air-conditioning system ofthe temperature and humidity control equipment is shown in Fig.6a.The water temperature at the outlet is–3 °C,and the temperature of the return water is 2°C.To allow for a safety margin in the refrigeration conditions,three coil refrigeration setups are used,with a refrigeration temperature of approximately 7°C for each coil refrigeration setup.Finally,it is easy to obtain an air flow temperature lower than 5°C,which is the dew point temperature,to ensure a relative humidity of less than 30%.

Integration of the diffractometer is performed by different motors and video systems.Four motor systems are applied:(1)the Kohzu motor system is used for Inline_camera_vert,Inline_camera_horz,Inline_camera_pitch,Inline_camera_yaw,and Inline_camera_focus;(2)the SmarAct motor system is used to adjust the position of the collimator and beamstop,and also sample_x,sample_y;(3)the Aerotech motor system is used for sample_z,gonio_phi,and gonio_vert;(4)the EPICS system is used for the air-bearing stages.To improve the stability of the motor systems,a Kohzu distributed hardware server(DHS),SmarAct DHS,and Aerotech DHS are developed using the C++programming language.Complex motion can be obtained by a combination of different DHSs or a combination of different DHSs with EPICS.For example,beam intensity adjustments are made by a combination of the Aerotech DHS,Kohzu DHS,and SmarAct DHS,and crystal centering is achieved by a combination of the Kohzu DHS and Aerotech DHS;scanning is obtained by calling a function of the Aerotech DHS.The video systems are divided into an inline camera video system,robot video system,and hutch video system.The online camera video system is used for crystal centering.The robot and hutch video systems are used for monitoring the status of the robot and hutch,respectively.

To integrate the EIGER detector into Blu-Ice,an EIGER DHS is developed using the C++programming language.All the functions of the EIGER detector are integrated into the EIGER DHS,and the EIGER DHS communicates directly with the EIGER detector control unit;therefore,the Blu-Ice system can control the EIGER detector through the EIGER DHS.

大学四年,金钻明用“绝不轻松”来形容。首任终身校长钱伟长将清华的教育理念带到上海大学,标准高、要求严、节奏快,本科教育实行短学期制、选课制和学分制,10周上课2周考试。物理基础课程很多,除了经典的“力热光电”物理课程外,还有原子物理和物理四大力学等,因此大学一二年级的学业比较重,三四年级上专业课时稍微轻松一点。由于兴趣加上把主要时间都用在学习上,金钻明说自己对物理越来越感兴趣,特别在毕业设计阶段越来越进入状态。

多媒体技术的滥用对阅读能力的培养毫无用处,造成适得其反的效果,然而适度合理的使用多媒体对提升小学生的阅读能力产生有效帮助。在阅读能力的培养中,尽量避免在阅读之前使用多媒体,以免干扰学生的阅读思维,让学生分心,导致在阅读中学生一直回想多媒体画面,对多媒体展示的内容心心念念。多媒体技术可以在阅读行为结束之后,作为老师分析内容的辅助来使用。老师为了便于学生理解文章,将文章结构进行梳理,用多媒体进行呈现,学生直观地了解了文章层次。这样有利于学生于学生学习老师的阅读思路,提高阅读效率和阅读能力。

Data can be collected from the EIGER detector using a single-image testing model;a testing image can be collected for initial screening of the quality of crystals.A C++program was developed to read HDF5 images,and an HDF5 diffraction image is automatically shown in Blu-Ice.Salt crystals and poorly diffracting crystals are not included in large-wedge dataset collection.The beam time and automatic data processing time can be signi ficantly reduced after initial screening using the single-image testing model.Moreover,data collection can be performed using a shutterless model for large-wedge(up to 3600°)data collection.A full dataset containing a master file and multiple HDF5 image files is obtained from the EIGER DHS.The packed HDF5 image number is automatically calculated by dividing the total data collection time by 20 s;therefore,each packed HDF5 image is generated from the EIGER detector with a 20-s serial interval.The advantages of using a 20-s serial interval for each packed HDF5 images are as follows:(1)Users do not need to set up the frame number in each packed HDF5 image file,which can signi ficantly save beam time;(2)a 20-s serial interval can provide the necessary time for HDF5 image processing to show in Blu-Ice for monitoring by users.

品牌不仅仅是一种符号结构,一种产品的象征,也是企业、产品、社会的文化形态的综合反映和体现;品牌是企业一项产权和消费者的认识,更是企业、产品与消费者之间关系的载体。乌村目前所使用的品牌是“乌镇”,但是乌村是独立于乌镇东西栅、古北水镇的一个新项目,所以一定要打造属于乌村自己的品牌,比如设计属于自己的logo以及代表乌村的吉祥物,争取成为继西栅和古北水镇以后一块金字招牌。

Fig.7 Structure of hardware control

4 Data collection

Compared to the original experimental setup,the current instruments offer more precise positioning during sample rotation and continuous readout from the detector.Owing to the features of the EIGER X 16M,such as the fast readout and the absence of readout noise,the ultra fine-φ slicing data collection strategy[7],which has proved to be capable of obtaining better data quality than fine-φ slicing[8],is recommended.For the EIGER X 16M,an optimal delta phi setting of 1/10 of the mosaicity is recommended,although this setting requires huge storage and powerful computational capabilities.

After equipment installation and control integration were completed,standard lysozyme samples were used to test the data quality of the beamline.The crystallization condition of the lysozyme was NaAc,pH 4.6,8%NaCl at 18°C.Crystals were cryoprotected with a cryoprotectant before being frozen in liquid nitrogen.The wavelength was 0.97918 A˚.The exposure parameters are a delta angle of 0.2°per frame and a 0.2 s exposure,and the total number of frames is 1800.Data were processed using the xia2 software suite.The diffraction image in Fig.8,which was obtained from the merged raw images,was offered for inspection by users.The crystallographic statistics are shown in Table 1.

外语学科的美国学学者,应该有全球视野,要充分认清文化的多样性和文明的多元化,充分展现人类命运共同体的包容与开放姿态。研究不是为了弘扬美国文化价值观,而是为了深度解读和认识美国,系统考察文化因素对相应美国历史与现实议题的作用与影响;同时,积极拓展外语学科内涵,让美国学成为新时代外语学科建设的重要内容,服务于国家对外战略需求,这是外语学科美国学学者的时代责任。

Fig.8(Color figure online)Diffraction image of lysozyme

Table 1 Crystallographic statistics of the lysozyme crystal

aValues in parentheses are those of the highest-resolution shell

Parameter Value Data collection Space group P43212 a(A˚) 78.35 b(A˚) 78.35 c(A˚) 36.97 α,β,γ(°) 90,90,90 Resolution(A˚) 36.97–1.35(1.39–1.35)a Rmeas(%) 6.0(127.5)a<I/σI> 17.8(1.7)a Completeness(%) 99.7(99.7)a Redundancy 12.2(12.2)a

5 Discussion and conclusion

The first self-integrated diffractometer,a temperature and humidity control facility,and the EIGER X 16M give the BL17U1 endstation a new look.The beamline is now in user operation after fast commissioning.Stable operation during user operation proves that the design of the highprecision goniometer is successful.Users’feedback on the new system will be recorded for use in developing the next version of the goniometer.

The EIGER detector not only speeds up data collection,but also makes new experimental schemes possible.The typical experimental setup parameters,such as the oscillation angle,exposure time,and beam attenuation,will be gradually transferred from the mode under the chargecoupled device detector to that under this EIGER detector.The huge amounts of raw data require greater storage capability and improved data analysismethodsand resources.The beamline staff will cooperate with the user community to exploit the new experimental schemes to facilitate users’projects.

References

1.R.L.Owen,J.Juanhuix,M.Fuchs,Current advances in synchrotron radiation Instrumentation for MX experiments.Arch.Biochem.Biophys.602,21–31(2016).https://doi.org/10.1016/j.abb.2016.03.021

2.Biosync,Biosync:a structure biologist’s guide to high energy data collection facilities(2017).http://biosync.rcsb.org.Accessed 1 Jan 1995

3.Q.S.Wang,F.Yu,S.Huang et al.,The macromolecular crystallography beamline of SSRF.Nucl.Sci.Tech.26,010102(2015).https://doi.org/10.13538/j.1001-8042/nst.26.010102

4.Q.Y.Pan,Q.S.Wang,Z.J.Wang et al.,An active beamstop for accurate measurement of high intensity X-ray beams.Nucl.Instrum.Methods Phys.Res.A 735,584–586(2014).https://doi.org/10.1016/j.nima.2013.10.011

5.H.Group,Support the DHF group (2017).https://support.hdfgroup.org.Accessed 18 Sep 1990

6.Q.Wang,S.Huang,B.Sun et al.,Control and data acquisition system for the macromolecular crystallography beamline of SSRF.Nucl.Tech.35,5–11(2012).(in Chinese)

7.A.Casanas,R.Warshamanage,A.D.Finke et al.,EIGER detector:application in macromolecular crystallography.Acta Crystallogr.D Struct.Biol.72,1036–1048(2016).https://doi.org/10.1107/S2059798316012304

8.M.Mueller,M.T.Wang,C.Schulze-Briese,Optimal fine φ-slicing for single-photon-counting pixel detectors.Acta Crystallogr.D Biol.Crystallogr.68,42–56(2012).https://doi.org/10.1107/S0907444911049833

Qi-Sheng Wang,Kun-Hao Zhang,Yin Cui,Zhi-Jun Wang,Qiang-Yan Pan,Ke Liu,Bo Sun,Huan Zhou,Min-Jun Li,Qin Xu,Chun-Yan Xu,Feng Yu,Jian-Hua He
《Nuclear Science and Techniques》2018年第5期文献

服务严谨可靠 7×14小时在线支持 支持宝特邀商家 不满意退款

本站非杂志社官网,上千家国家级期刊、省级期刊、北大核心、南大核心、专业的职称论文发表网站。
职称论文发表、杂志论文发表、期刊征稿、期刊投稿,论文发表指导正规机构。是您首选最可靠,最快速的期刊论文发表网站。
免责声明:本网站部分资源、信息来源于网络,完全免费共享,仅供学习和研究使用,版权和著作权归原作者所有
如有不愿意被转载的情况,请通知我们删除已转载的信息 粤ICP备2023046998号