一种新型钢卷尺零位标定方法

更新时间:2016-07-05

0 Introduction

According to the national standard steel tape verification regulation of JJG 4-2015[1], the metrological verification units verify half and integer meter scales of steel tape periodically. Normally, the indication errors of steel tape under test and standard tape can be calibrated by deviation of scales[2]. However, there are some special steel tapes with movable structures on zero position, as shown in Fig.1. Traditional zero scale collimation methods are not applicable to this kind of steel tape due to the lack of zero scale.

Fig.1 Steel tape with movable hook structure

The steel tape with movable hook structure on zero position is a widely used length measuring instrument. Push-type measurement and hook-type measurement are two different operation modes, and the zero position depends on frontal datum plane and rear datum plane of the hook, respectively, as shown in Fig.2. Displacement d is equal to the thickness of the hook (including cock quantity), which compensates for the deviation between the two datum planes.

Fig.2 Principle of hook structure

Here e0 is the indication error of measured steel tape of 500 mm, and it is regarded as the zero error in conventional verification of the whole steel tape[3]. On the long sized stage of whole steel tape, e0 will affect ei,the indication error of subsequent verification, and ei can be calculated by

ei=Ni-Li+e0,

(1)

where Ni is indication value, and Li is standard value. Δ will be introduced in the process of zero scale collimation. Δ is caused by the deviation of zero scales which should have been aligned completely.

As previously mentioned, it is necessary to develop a new zero calibration method for the steel tape. In this paper, a new system is proposed. The zero position of hook structure is determined by standard datum planes, and indication error of 500 mm scale is calibrated by standard scales. Taking Δ′, the indication error of 500 mm scale, as the compensation value of e0 while zero position is determined. Supposing that the revised indication error of 500 mm scale is and the revised indication error of subsequent calibration is The mathematical expressions are

AutoLISP可以利用命令“getfiled”打开文本文档[2],因此,将程序保存在文本文档里,在AutoCAD的命令栏里输入“lathe”命令,程序打开一个对话框供用户选择程序,如图5所示。

4.事业单位提取职工福利基金比例超过非财政补助结余的40%。这样提取不符合财政部《关于事业单位提取专用基金比例问题的通知》的规定。究其原因,主要是会计人员不熟悉事业单位制度规定造成的。

(3)

1 Analyses of existing methods

To meet the needs of zero scale collimation in steel tape calibration, a variety of methods are researched.

1.1 Zero adjustment system with manual mechanism

The operation interface is designed as shown in Fig.7. In addition to 500 mm scale, other integer decimeter scales between 0 mm and 500 mm can also be calibrated by this system. The image in the left side is a processed binary image. The indication error can be calculated according to pixel equivalent. Identifying line center and horizontal projection of the pixels in binary image[9] are used in image processing. What’s more, the mode of operation, push-type or hook-type, can be chosen before calibration, so does the type of measured tape.

Fig.3 Zero adjustment system with manual mechanism

The zero adjustment system was composed of loading mechanism, zero scale collimation mechanism, manual adjustment mechanism and computer vision system. The relative position between zero scale and standard scale of measured steel tape was shown by the image signal. Then steel tape’s conveying was driven by a manual adjustment mechanism, and zero scale collimation was realized.

保利滨湖广场裙楼部分共三层,公共区域包括大堂、商务、餐饮和康体四部分功能,均整合在其中。裙楼西侧面对文翰湖,为主要景观面,东侧为凹入式内景庭院,南侧设计为婚宴花园,引景入室,处处有景。酒店大堂主入口设在西侧,步入大堂,正对的大堂吧两层通高,园林景观一览无余,接待前台、全日制餐厅和特色餐厅设在大堂左侧,亦共享庭院景观。二层设有行政餐厅和中餐厅,满足不同商务人群的饮食选择。

This method has a better accuracy compared with optical reading devices like microscopes[5]. However, it has some limitations. Firstly, due to manual adjustment, it is difficult to guarantee the stability. What’s more, the two operation modes of steel tape are not taken into account.

1.2 Zero collimation system with motorized platform

Lei T[6] mentioned an automatic zero collimation system, as shown in Fig.4. After the rough localization, zero scale was in the scope about ±2 mm from the standard datum line. Then the position signal would be sended to computer, and the distance could be calculated. Next conveying motor drove the micro-displacement platform with measured steel tape to move to the standard datum line. At last servo motor was stopped while the zero scale and standard datum line were coincided.

Fig.4 Zero collimation system with real-time motorized platform

Automatic conveying and collimation can save labor. As to a servo system, a very high degree of accuracy is required when driving the steel tape movement. But it is nearly impossible to drive micro-displacement platform to align zero position without any deviation. Besides, this system is not applicable to the steel tape with movable hook structure.

2 New zero calibration system

The zero calibration method proposed in this paper is specifically designed for the steel tape with movable hook structure. This system is aimed to solve the indeterminacy of zero position. Standard datum planes and standard scales are designed to determine zero position and calibrate indication error. A new system based on this principle is put forward, as shown in Fig.5.

Fig.5 Mechanical structure of steel tape zero calibration system

2.1 Main component elements

This system adopts a float equivalent method[8]. During the image processing, the two standard scales in the field, whose distance is known exactly, are used as standard to calculate dynamic pixel equivalent. As previously mentioned, distance between the two standard scales on board is l=4 mm. The corresponding pixel number n can be counted in image processing. The pixel equivalent b can be calculated by

以上几首引用了《三国演义》里面的典故。例1、2是女主人向男方同伴敬酒时唱的歌,一首歌,一杯酒。这里用到了《三国演义》中“五虎上将”、“桃园结义”等典故。这里的大体意思为女方出嫁,是有人要把她“抢走”,所以希望来唱歌的同伴们“保护”她。歌词里女方表达了希望男方像关羽那样勇猛,能保得住荆州(实际指出嫁姑娘),而男方则用“桃园三结义”表达与女方之间有深情厚谊,希望女方别抛弃他。

Table 1 Experimental devices and key parameters

ExperimentaldeviceKeyparameterCMOSindustrialcameraModel:MV⁃EM500C/MResolution:2592×1944Pixelsize:2.2μm×2.2μmOpticallensModel:M1224⁃MPW2Focallength:12mmLightsourceCoaxialparallellightsourceVerificationplatformWorkinglength:500mmLinearguiderailSBI15SV⁃2⁃K1⁃830N

2.2 Principle

In this new system based on machine vision, an indirect approach is proposed. The calculation of deviation value is done by camera and computer, which can get much more accuracy results than traditional observational methods. The distance between standard datum planes and standard scales is regarded as a datum quantity. We compared this datum quantity with the distance between zero position and 500 mm scale, it can be found that the indication error is equal to the zero error. Push-type measurement datum and hook-type measurement datum are designed to get the measured value. Then the result can be got indirectly, although zero position is a plane rather than a scale. Traceability operation is designed to confirm the accuracy of datum quantity.

2.3 Calibration operation

Steps of the zero calibration are as follows.

1) Determine the type of measured steel tape and operation modes.

2) Push or catch the hook structure tightly against the positioning blocks and fix the box of tape.

3) Camera collects images of 500 mm scale.

4) Computer processes the images to calculate the pixel equivalent and takes the equivalent as standard component, and then the indication error is calculated.

5) Repeat steps 2 and 3 while changing the operation mode.

2.4 Traceability operation

Standard datum planes are determined by traceability operation. The mechanical structure is composed of positioning blocks, loading mechanism and micrometer handwheel. The positioning blocks are taken as reference of different standard datum planes of push-type and hook-type measurements. Combined with different modes of operation, steps of the traceability are as follows.

第一,医疗水平总体偏低。海南医疗旅游产业发展最大的短板就是高端医疗机构和医疗人才匮乏,前沿医疗技术发展水平偏低。目前,海南仅有2家私立医疗机构通过JCI认证,分别是海南现代妇婴医院和海南瑞韩医学美容,难以满足高端国际医疗旅游的需求。

按照条例,哈尔滨市水行政主管部门应当会同市发展改革、环境保护城乡规划、农业、林业等部门,根据市水生态保护与修复规划,组织编制市水生态监测规划,报市人民政府批准。市和区、县(市)水行政主管部门应当根据市水生态监测规划,组织设立水生态监测站点。

Fig.6 Mechanical structure of standard datum location device

2) Determine the position of standard datum planes in hook-type measurement. After the standard datum in push-type has been fixed, the 4 mm gauge block[7] is sandwiched between two positioning blocks as a higher precision standard component. We should ensure that the distance between the two datum planes is equal to the distance between the two standard scales on board.

1) Determine the position of standard datum planes in push-type measurement. The standard component used as a higher precision reference is knife-type steel scale, which is designed by Tianjin Institute of Metrological Supervision and Testing. The knife-type steel scale has a standard length of 500 mm between standard scale line and standard surface. Micrometer handwheel and setscrew are used to determine and fix the the position of push-type standard datum planes. It is shown in Fig.6.

2.5 Float pixel equivalent method

Pixel equivalent is an important parameter to describe the relationship between pixel of digital image and actual size of measured piece in the 2D image measuring system. Also it plays a role of precision standard in machine vision system.

Zero calibration system is composed of mechanical desktop, location structure of standard datum planes, loading mechanism, high precision dovetail guide rails, one-dimensional displacement mechanism,multi-freedom fine-tuning structure of camera, machine vision system and computer system software. Experimental devices and key parameters are shown in Table 1.

随着乡村振兴战略的推进,农村正在巨变,不仅仅是公路四通八达,民居整洁漂亮,生态愈发优美,百业更加兴旺,而且农村中各种新生事物也正在为人们津津乐道。但是,如果说到短板,许多农民弟兄和基层干部也有连连摇头的时候,不在物质生活,而在精神领域,特别是文明乡风、伦理道德有堪忧之处,有些地方说是“短板”,毋宁说是个“重灾区”。

(4)

In this system, the width of field is around 40 mm, and effective pixel of industrial camera is 2 592×1 944. It is obvious that the calibration process has a high degree of accuracy, and the resolution is 0.02 mm.

2.实习时间安排不当。酒店管理专业的学生实习时间一般为三个月到一年,长时间的实习会使学生产生一定的厌倦心理,严重影响了学生对于本行业的认识,甚至会使部分学生在毕业后出现改行的情况。

“要改变自己的命运就要有不服输的精神,敢闯敢拼才会赢”。2010年,团场出台了一系列优惠政策,鼓励职工发展自营养殖,他辞职下海当年养鸡12000只,成活率达到90%以上,种植的100亩棉花单产走在全团前列,当年实现纯收入49万元。他在团部购买了楼房,买了台754播种机,为了儿子上学方便,买了小汽车。当年机车仅拉棉花和打杆两项,实现纯收入7.7万元,种植棉花和养殖鸡纯收入达35万元。

2.6 Operation and displace interface

Yu C L[4] used a direct method to align zero scale, as shown in Fig.3.

Fig.7 Operation and displace interface

3 Experiments and analysis

An experiment was designed to examine repeatability of new system. Repeatability tests were carried out at 500 mm scale on the same steel tape. Indication errors were recorded in two different operation modes. The records are shown in Table 2 and Fig.8.

Table 2 Experimental results

No.Hook⁃typemeasurementPush⁃typemeasurementb(μm)n′e′0(mm)b(μm)n′e′0(mm)118.774.734680.0888918.759.276980.17392218.784.776790.0897018.779.370500.17591318.784.814460.0904218.779.291370.17436418.774.905740.0921018.749.259020.17353518.784.942310.0928018.779.303960.17460618.774.775390.0896518.749.241010.17314718.774.867760.0913918.759.250000.17348818.784.937010.0926918.759.248200.17345918.784.616260.0866918.779.251800.173631018.784.808370.0903018.769.213020.17283Max18.784.942310.0928018.779.370500.17591Min18.774.616260.0866918.749.213020.17283Range0.010.326500.006110.030.157480.00308Mean18.7764.817870.0904618.7579.270580.17388U0.001300.00053b:Thepixelequivalent;n′:Thepixelnumberofindicationerror;e′0:In⁃dicationerror;U:Uncertainty

Fig.8 Records of experiment

The mean value can be calculated by

这种病例病程较长,早期表现食欲不振,日渐消瘦,结膜苍白,被毛粗乱,粪便干燥而少,体温正常,但也有的体温稍下降,严重者卧地不起。

(5)

where n is the times of calibration. If we take as the calibration result, its standard uncertainty uA is

(7)

Taking uncertainty of measurement U as expanded uncertainty, it is

U=kuA,

(8)

where k is coverage factor, generally k=2.

As shown in Table 2 and Fig.8, the system’s calibration repeatability is approximately 0.006 mm, and the uncertainty is less than 0.002 mm, which is sufficient for practical use. Besides, the range of pixel equivalent is very little, which can be neglected. The traditional observational system usually introduces errors caused by the naked-eye and zero scale alignment process, and the repeatability is around 0.1 mm. It is obvious that this novel zero calibration method aiming to solve indeterminacy of zero position is far superior to the traditional methods in not only accuracy but also the repeatability of the result.

4 Conclusion

In this paper, a new zero calibration method of steel tape is put forward to deal with the indeterminacy of zero position caused by movable hook structure. Standard datum planes and standard scales are designed as the reference standard. The float pixel equivalent method is applied to ensure the accuracy of calibration process. Meanwhile, cooperated with 5 m verification platform[10-12], this system shows obvious superiority over traditional methods in both accuracy and stability.

References

[1] China JJG 4-2015. Institute of metrological supervision and testing. Tianjin: Tianjin Institute of Metrological Supervision and Testing, 2015.

[2] Gou J S. Research of the new steel tape automatic verification mechanism. Tianjin: Tianjin University, 2015.

[3] Chen S T, Qian X Y, Zhou M. Research of automation steel tape measurement system based on PLC and Computer Vision technology. In: Proceedings of IEEE World Congress on Software Engineering, 2009, (1): 251-255.

[4] Yu C L, Zang Y G, Xie W H. New steel tape measure verification equipment research. China Metrology, 2012, 56(5): 76-77.

[5] Li S H. Verification zero deviation of sounding steel tape. Measurement Technique, 1999, (4): 27-28.

[6] Lei T, Zeng B C. Steel tape automatic zero alignment system based on Labview. Metrology & Measurement Technique, 2016, 43(4): 57-59.

[7] Huang L H, Liu N. Length value in-kind transfer standard-gauge block. Brand & Standardization, 2010, 12(4): 41-43.

[8] Gou J S, Wang Z, Shen X L. New float equivalent calibration method for 2D image measuring system. In: Proceedings of SPIE International Conference on Optical Instruments and Technology, 2015, (9623): 962310-962318.

[9] Zhao B Y, Wei C Q, Kang H F. Image binarization method of steel tape based on edge detection. Modern Manufacturing Engineering, 2012,15(12): 101-104.

[10] Mao Z H, Du Z P, Zhang X. Low cost and high precision steel tape automatic calibration device. China Metrology, 2011, 22(2): 88-90.

[11] Wei C Q, Wang X G, Kang H F. Steel tape automatic verification system based on image processing. Instrument Technique and Sensor, 2012, (6): 88-90.

[12] Xu J. Steel tape automatic calibration device based on machine vision detection technology. Measurement Technique, 2009, (11): 50-53.

李一鸣, 王 仲, 陈 曦, 路瑞军 (天津大学 精密测试技术及仪器国家重点实验室, 天津 300072)
《Journal of Measurement Science and Instrumentation》2018年第2期文献

服务严谨可靠 7×14小时在线支持 支持宝特邀商家 不满意退款

本站非杂志社官网,上千家国家级期刊、省级期刊、北大核心、南大核心、专业的职称论文发表网站。
职称论文发表、杂志论文发表、期刊征稿、期刊投稿,论文发表指导正规机构。是您首选最可靠,最快速的期刊论文发表网站。
免责声明:本网站部分资源、信息来源于网络,完全免费共享,仅供学习和研究使用,版权和著作权归原作者所有
如有不愿意被转载的情况,请通知我们删除已转载的信息 粤ICP备2023046998号