A comparative study of TR Band and a new hemostatic compression device after transradial coronary catheterization

更新时间:2016-07-05

INTRODUCTION

The traditional site of vascular access for coronary angiography or angioplasty has been through the femoral artery, and the vast majority of coronary procedures were performed this way previously (1,2).However, this can also potentially cause vascular complications, such as pseudoaneurysm,arteriovenous fistula, and most seriously,retroperitoneal bleeding (3). In 1989, Campeau (4)introduced transradial access (TRA) for performing coronary angiography, and in 1993, Kiemeneij and Laarman (5,6) reported his experience with coronary angioplasty through the radial route. The transradial route is increasingly popular, with the primary advantages of allowing earlier mobilization of patients post procedure and significantly less vascular complications compared with transfemoral access(7,8).

As the arterial sheath can be removed immediately after the procedure, with mechanical compression applied over the puncture site of the radial artery, early hemostasis can be achieved. This allows early mobilization and discharge from hospital (9). Several types of hemostatic compression devices have been used, and proved effective and safe. The TR Band is produced by Terumo in Japan, and is now commonly used in China to assist in hemostasis of the radial artery after a transradial procedure. By injecting a certain volume of air into the transparent balloon in the TR Band, the radial artery is compressed and hemostasis is achieved (10,11). However, because the wrist size of patients varies, the fixed volume of air may produce differing pressures on wrists, resulting in unstable efficacy of compression. A newly designed radial compression device that injects air to the balloon with a particular pressure may supply stable and appropriate compression, and seems to be promising in clinical practice. The objective of our study was to compare the effect of this new hemostatic compression device with the TR Band on efficacy of hemostasis, patient comfort, local vascular dysfunction, and radial artery occlusion.

MATERIALS AND METHODS

Study subjects

The enrolled patients were hospitalized at the cardiology department in Shanghai Rui Jin Hospital during June and July 2018. Patients with suspected or established stable coronary artery disease were considered for coronary angiography and intervention through the transradial route. Exclusion criteria included acute infectious disease, acute myocardial infarction, uremia, mental disorder, thrombocytopenia,hemorrhagic disease, forearm arteriovenous fistula, or history of previous ipsilateral transradial procedure.Each patient received an Allen’s test, but was not excluded on the basis of an unfavorable result. The study complied with the Declaration of Helsinki and was approved by the ethics committee of Shanghai Jiao Tong University. Informed consent was obtained from the patients prior to enrollment.

Study design

For this prospective randomized study the patients were divided randomly into two groups according to radial compression device before procedure: a TR Band group (Terumo, Japan) and a group receiving the new hemostatic compression device. Randomization was performed using computer-generated random numbers. All procedures were performed by the same two interventional cardiologists at a single cardiac catheterization laboratory under similar operating conditions. Radial hemostatic compression devices were applied following coronary catheterization.During and following the hemostatic compression, the outcomes were recorded.

Study outcome

The outcome evaluation included efficacy of hemostasis, patient comfort, local vascular dysfunction, and radial artery occlusion (RAO).Efficacy of hemostasis was assessed by whether patients had errhysis or hematoma during and following the hemostatic compression. Patient comfort was evaluated by whether the patients had pain or a feeling of numbness around the puncture site during hemostatic compression. Pain levels were graded as no pain, mild pain, or moderate to severe pain. Feeling of numbness was graded as no numbness,mild numbness, or moderate to severe numbness.Local vascular dysfunction was judged by pulse loss in the distal artery to the hemostatic compression device or obstruction of venous reflux during hemostatic compression. RAO at discharge was confirmed by the absence of palpable radial artery pulsation and one of the following two verification tests: abnormal reverse Allen’s test or absent Doppler flow signal on hand-held Doppler.

Clinical data collection

Various types of radial hemostatic compression device have been used following percutaneous coronary catheterization and have proved effective,safe, and well tolerated. Previous studies have compared efficacy of hemostasis, patient comfort, and local vascular complications among hemostatic compression device types (9-11). Through its transparent structure designed for visual control and selective compression of the radial artery to allow blood return and preserve patency, the TR Band assists in maintaining radial artery patency at the time of hemostasis in order to prevent future RAO and canoften produce optimal patient comfort (10,11). A previous comparative study indicated that the TR Band performs well compared with other hemostatic compression devices (10). A new type of radial compression device designed to inject air at a particular pressure has been produced, with the intention of providing stable and appropriate compression on the puncture site. To our knowledge,this is the first randomized trial evaluating the efficacy and tolerance of this new radial compression device by comparing it with the commonly used TR Band.

Procedures

The radial artery was punctured with a 21-gauge arterial needle before transradial coronary catheterization was performed. The needle was then withdrawn and a 6F introducer sheath inserted through the guidewire. All introducer sheath kits used were from the same manufacturer (Terumo, Japan). The remaining procedures were performed according to the operator’s preference. All patients received intravenous heparin after sheath insertion, using 70 U/kg for diagnostic procedures. Additional heparin was given for percutaneous coronary intervention(PCI) if needed.

Hemostatic device application

The introducer sheath was removed when the procedure was finished, and hemostasis compression was simultaneously applied using either TR Band(Figure 1A) or the new hemostatic compression device (Figure 1B), according to randomization. The TR Band was applied according to the manufacturer’s instructions. A volume of 14-16 mL of air was injected slowly to the balloon according to the patient’s wrist size, while simultaneously removing the sheath. When the air was fully inserted, the sheath was completely removed. Reduction in compressive pressure was started 4 hours after the procedure;thereafter operators removed 2 mL air every 2 hours until hemostasis was achieved, and the device was then removed. Similarly, the new hemostatic compression device (Shanghai KDL medical instrument company, China) was used according to the manufacturer’s instructions. Air was injected slowly to the balloon until a pressure of 250 mmHg was achieved, guided by a pressure indicator. When the air was fully inserted, the sheath was removed.Reduction in compressive pressure was started 4 hours after the procedure. Operators subsequently removed 1 mL of air every 2 hours until hemostasis was achieved, and the device was then removed.

正中神经、尺神经SCV、MCV与自主神经病变存在广泛相关性,差异有统计学意义(P<0.05)。正中神经与心脏自主神经功能关系更为密切,SCV与心脏自主神经功能关系更为密切。见表2。

Figure 1. (A) TR Band. The band and injector are included. (B) New hemostatic compression device. The injector, pressure indicator, and band are included.

Figure 2. Air pressure in balloon and actual pressure on artery according to balloon air volume.

Before clinical application of the new hemostatic compression device, a series of experiments was performed to test the efficacy of hemostatic compression. From injection of air into the balloon in the band, air pressure in the balloon was produced followed by pressure on the artery. To study air pressure changes in the balloon and actual pressure on the artery following the use of different air volumes in the balloon, an experiment was conducted, with results data shown in Figure 2. An understanding of the relationships between balloon air volume, balloon air pressure, and actual pressure on the artery assisted in the design of the pressure indicator and the new hemostatic compression device.

Statistical analysis

Age, gender, height, weight, BMI, blood pressure,hypertension, lipid disorders, and current smoking prevalence were similar in patients with and without RAO (Table 3). Patients with RAO at discharge received less heparin during percutaneous coronary catheterization compared with those without RAO at discharge (P = 0.014), and fewer patients receiving the new hemostatic compression device following percutaneous coronary catheterization experienced RAO (11.1% vs. 53.6%, P = 0.015). Diabetes mellitus prevalence and prevalence of patients receiving PCI was higher in patients with RAO at discharge,although the difference was not significant (P = 0.068,and P = 0.074 respectively).

Obstruction of venous reflux was experienced by a total proportion of 16.1% of the patients. Significantly fewer patients in the new hemostatic compression device group experienced obstruction of venous reflux than in the TR Band group (6.8% vs. 25.4%, P =0.006).

RESULTS

Baseline Characteristics

The proportion of patients experiencing pulse loss in the artery distal to the new hemostatic compression device was 13.6% (Table 2). There were fewer patients in the new hemostatic compression device group experienced pulse loss than in the TR Band group (5.1% vs. 22.0%, P = 0.007).

Efficacy of hemostasis

The total proportion of patients developing errhysis or hematoma during and following hemostatic compression was 12.7% (Table 2), and was not significantly different between the two groups (13.6%vs. 11.9%, P = 0.782).

Patient comfort during hemostatic compression

Pain levels differed significantly between the two groups (Figure 3; P = 0.003), with more patients in the new hemostatic compression device group experiencing no pain than in the TR Band group (40.7%vs. 33.9%). Less participants in the new hemostatic compression device group experienced moderate to severe pain than in the TR Band group (1.7% vs.22.0%).

Table 1 Clinical characteristics of study patients according to hemostatic compression device

BMI: body mass index; BP: blood pressure; PCI: percutaneous transluminal coronary intervention. *Values are mean ± standard deviation, median (interquartile range), or n (%).

Variable* All (n=118) TR Band (n=59) New hemostatic compression device(n=59)P value Age (years) 64.5±10.4 64.5±9.73 64.5±11.1 0.965 Men (n, %) 79 (66.9%) 40 (67.8%) 39 (66.1%) 0.845 Height (m) 1.66±10.2 1.68±7.92 1.65±11.9 0.162 Weight (kg) 69.5±12.2 71.5±12.0 67.6±12.3 0.091 BMI (kg/m2) 25.3±5.09 25.5±3.98 25.1±6.03 0.727 Heparin dose (U) 3000 (3000-7000) 5000 (3000-7000) 3000 (3000-7000) 0.150 Systolic BP (mmHg) 131.6±23.1 133.8±23.8 129.4±22.5 0.310 Diastolic BP (mmHg) 75.2±9.46 75.7±9.82 74.8±9.15 0.608 Current smoking (n, %) 10 (8.5%) 5 (8.5%) 5(8.5%) 1.000 Diabetes mellitus (n, %) 34(29.1%) 19(32.2%) 15 (25.9%) 0.450 Hypertension (n, %) 79 (66.9%) 44 (74.6%) 35 (59.3%) 0.078 Lipid disorders (n, %) 43 (36.4%) 21 (35.6%) 22 (37.3%) 0.848 PCI (n, %) 46 (39.0%) 25 (42.4%) 21 (35.6%) 0.450

Table 2 Local vascular outcome during and following the application of hemostatic compression devices

RAO: Radial artery occlusion.

Variable All (n=118) TR Band (n=59) New hemostatic compression device(n=59)P value Errhysis or hematoma (n, %) 15 (12.7%) 8 (13.6%) 7 (11.9%) 0.782 16 (13.6%) 13 (22.0%) 3 (5.1%) 0.007 Obstruction of venous reflux (n, %) 19 (16.1%) 15 (25.4%) 4 (6.8%) 0.006 RAO at discharge (n, %) 9 (7.6%) 8 (13.6%) 1 (1.7%) 0.015 Loss of pulse in artery distal to hemostatic compression device (n, %)

Figure 3. Patient discomfort during application of the hemostatic compression devices. A: Comparison of pain levels in patients with the new hemostatic compression device and the TR Band (P = 0.003). B: Comparison of feeling of numbness in patients with new hemostatic compression device applied and those with the TR Band (P = 0.010).

Table 3 Clinical characteristics of study patients with or without RAO at discharge

RAO: Radial artery occlusion; BMI: body mass index; BP: blood pressure; PCI: percutaneous transluminal coronary intervention. Values are mean ± standard deviation, median (interquartile range), or n (%).

Variable All (n=118) With RAO (n=9) Without RAO (n=109) P value Age (years) 64.5±10.4 64.4±7.32 64.5±10.6 0.987 Men (n, %) 79 (66.9%) 6 (66.7%) 73 (67.0%) 0.985 Height (m) 1.66±10.2 1.68±8.25 1.66±10.3 0.571 Weight (kg) 69.5±12.2 72.9±15.2 69.3±12.0 0.395 BMI (kg/m2) 25.3±5.09 25.2±5.07 25.9±5.65 0.719 Heparin dosage (U) 3000 (3000-7000) 3000 (3000-3000) 5000 (3000-7000) 0.014 Systolic BP (mmHg) 131.6±23.1 128.2±20.1 131.9±23.4 0.650 Diastolic BP (mmHg) 75.2±9.46 77.7±12.2 75.0±9.23 0.420 Current smoking (n, %) 10 (8.5%) 0 (0.0%) 10 (9.2%) 0.342 Diabetes mellitus (n, %) 34 (29.1%) 5 (55.6%) 29 (26.9%) 0.068 Hypertension (n, %) 79 (66.9%) 8 (88.9%) 71 (65.1%) 0.145 Lipid disorders (n, %) 43 (36.4%) 4 (44.4%) 39 (35.8%) 0.604 PCI (n, %) 46 (39.0%) 1 (11.1%) 45 (41.3%) 0.074 New hemostatic compression device(n, %)59 (50.0%) 1 (11.1%) 58 (53.2%) 0.015

Numbness level during radial artery compression was significantly different between groups (Figure 3;P = 0.010). More patients experienced no numbness in the new hemostatic compression device group compared with the TR Band group (40.7% vs. 32.2%).Moderate to severe numbness was reported less often in the new hemostatic compression device group than in the TR Band group (1.7% vs. 18.6%).

Local vascular dysfunction during hemostatic compression

A total of 133 consecutive hospitalized patients were screened and 118 patients were finally included in the study. Fifteen patients were excluded because they had undergone a previous ipsilateral transradial procedure. The average age was 64.5 ± 10.4 years.Seventy-nine patients (66.9%) were male. The demographic and clinical features of the patients,according to radial compression device, are shown in Table 1. The data suggested that patient age, gender,height, weight, body mass index, blood pressure,heparin dosage, hypertension, prevalence of diabetes mellitus and lipid disorders, current smoking status,and prevalence of PCI were similar in both groups.PCI prevalence was 39.0% (46 patients) and the other 72 patients received only coronary angiography.

可以看出,我们将学生自主学习的任务放在了线上,给学生更高的自主权,让学生可以更加有选择性的,在老师给予的大量学习材料中,根据自己的知识掌握程度进行学习。线下的时间主要用在师生互动上。因为有了线上学习的基础和反馈,线下交流会更加有针对性,也更有效果。

(2) 在以0°(垂直线路)方向输入地震波时.钢管塔的位移和应力将达到最大值.最大位移出现在地线挂点处,最大应力出现在塔头斜材位置处.

Incidence of radial artery occlusion at discharge

RAO at discharge was experienced by 7.6% of the patient total (Table 2), with fewer patients in the new hemostatic compression device group experiencing RAO than in the TR Band group (1.7% vs. 13.6%,P=0.015).

并不是所有撒料区域都符合需要撒料的条件.例如,若某些区域面积过小,可能会导致装甑撒料执行机构最小撒料量大于实际所需撒料量,故可以根据分割完成的撒料区域所拥有的像素点数量对撒料区域做进一步的筛选,并将撒料区域像素点的数量作为撒料区域面积.图7是撒料区域面积大于的撒料区域,并保留图6中的编号.

Clinical characteristics associated with RAO at discharge

Data were analyzed using SPSS 22.0 (SPSS Inc.,Chicago, IL, USA) software. Continuous variables with a normal distribution were expressed as mean ±standard deviation, and variables with a skewed distribution were reported as median (interquartile range). Categorical variables were expressed as frequency and percentage. The chi-square test was used to compare categorical variables between several groups. The independent-sample t test or Mann-Whitney U test were used to compare continuous variables with normal or skewed distribution between two groups, respectively.Multiple logistic regression analysis was used to assess the independence of the association between risk factors and RAO. The odds ratio (OR) and 95%confidence interval (CI) were calculated. Two-sided P< 0.05 was considered significant.

Many studies have proved that the TR Band can efficiently assist in achieving hemostasis. The incidence of bleeding, including errhysis or hematoma,during application of TR Band radial compression was 14.2%-26.3% in previous studies (10,12,13). Our study indicated that 13.6% of patients developed errhysis or hematoma with application of the TR Band;a similar proportion of patients were detected with errhysis or hematoma (11.9%) following application of the new hemostatic compression device. Our study data suggested that both of these hemostatic compression devices were effective in achieving hemostasis.

Table 4 Independent risk factors associated with RAO at discharge

CI: confidence interval. PCI: percutaneous transluminal coronary intervention. *Multiple logistic regression analysis was used to calculate the odds ratio (OR) of various risk factors associated with RAO (radial artery occlusion).

Variable OR* 95% CI P value Age (per year) 0.981 0.887-1.084 0.702 Men (1=no, 2=yes) 1.055 0.182-6.105 0.953 Heparin dosage (per U) 0.998 0.996-1.000 0.034 Diabetes mellitus (1=no, 2=yes) 7.122 1.225-41.411 0.029 PCI (1=no, 2=yes) 36.632 0.425-3154.345 0.113 New hemostatic compression device (1=no, 2=yes) 0.062 0.006-0.675 0.022

DISCUSSION

Clinical characteristics were obtained by hospital chart review and patient interview. Body mass index(BMI) was calculated as body weight in kilograms divided by body height in meters squared (kg/m2).Current smoking was defined as smoking currently and smoking more than one cigarette daily for at least one year continuously. Patients with total cholesterol≥ 5.7 mmol/L, or low-density lipoprotein cholesterol ≥3.6 mmol/L, or triglyceride ≥ 1.7 mmol/L, or receiving treatment with antihyperlipidemic agents due to hyperlipidemia were considered as having a history of lipid disorder. Hypertension was diagnosed as systolic blood pressure (SBP) ≥ 140 mmHg, or diastolic blood pressure (DBP) ≥ 90 mmHg, or when patients were being actively treated with anti-hypertension drugs.Diabetes mellitus was diagnosed by a fasting plasma glucose test showing ≥ 7.0 mmol/L, or by a random plasma glucose test showing ≥ 11.1 mmol/L, or when patients were actively receiving therapy using insulin or oral medications for diabetes.

Multiple regression analysis was used to confirm the independent risk factors associated with RAO at discharge. Basic demographic data such as age and gender were entered into the logistic regression equation. At the same time, discrepant risk factors (P< 0.10) between two groups were screened using univariate analysis and also entered into the multiple regression analysis. Lower heparin dosage (OR: 0.998,95% CI: 0.996-1.000, P = 0.034) and diabetes mellitus(OR: 7.122, 95% CI: 1.225-41.411, P = 0.029) were independently associated with RAO at discharge(Table 4). However, application of the new hemostatic compression device was independently associated with low incidence of RAO at discharge (OR: 0.062,95% CI: 0.006-0.675, P = 0.022).

In our study, the new hemostatic compression device supplied excellent patient comfort, with more patients experiencing no pain or feelings of numbness in the new hemostatic compression device group,compared with the TR Band. Improvements to patient comfort such as this are beneficial in alleviating suffering. Although in the application of the TR Band 14-16 mL of air is injected into the balloon according to the size of the patient’s wrist, the pressure on the puncture site fluctuates, varying widely among patients. Thus, the band may be tight in some patients and loose on others. This instability condition may cause patient discomfort. However, with the new hemostatic compression device, a specific pressure of air is injected into the balloon for a precise fit and optimal patient comfort.

Pulse loss in the artery distal to the hemostatic compression device and obstruction of venous reflux was significantly lower with the new device. A volume of air that is virtually fixed is injected into the balloon when applying the TR Band, and thus the band may be very tight if the band is tied too tightly or the patient’s wrist is big. As a result, too much compression on the radial artery may cause loss of arterial pulse distal to the hemostatic compression device. Simultaneous obstruction of venous reflux also tended to occur. Too much compression on the artery and vein may cause patient discomfort and increase the risk of local vascular complications (14).

由此可见,在一段成熟的感情中,恋人们应该做到相互了解,彼此尊重,对任何信息的隐瞒和逃避都有可能造成误解并导致爱情的失败。另外,对于已经犯了的错误,人们不应该用更多的谎言去遮掩,而是要勇敢的承认和面对。雷与戴安娜的感情正是因为一个个叠加的谎言而走向失败。相反,汤米反思了自己的过失后,勇敢地告诉吉娜和丹尼他的经历并获得了他们的理解,三人的关系因此而得到修复。

RAO is one of the few postprocedural complications of TRA for diagnostic and interventional procedures. Previous studies have reported RAO rates ranging from 3% to 12% (7,12,15).Data in our study suggested that a lower heparin dose and diabetes mellitus were independently associated with RAO at discharge. These results were consistent with the results of a previous study, which also demonstrated that diabetes mellitus was independently associated with RAO (16). Several studies have shown that heparin use helps to reduce the occurrence of RAO (10,14,16). Heparinization, when safe, is important and probably aids recanalization after an occlusive hold, by making the local environment less thrombotic.

In our study, in contrast to the TR Band, the application of the new hemostatic compression device was independently associated with lower incidence of RAO at discharge, with adjustment for potential confounders. Excessive compression on the radial artery tended to be avoided because a fixed pressure,but not fixed volume, of air was injected into the balloon when the device was applied. In fact, a previous study proved that use of pneumatic compression guided by mean blood pressure can significantly lower the incidence of RAO (12). The related mechanism is similar.

一方水土养一方人,作为榆盘人,年年岁岁种洋芋,岁岁年年吃洋芋,再轻轻地哼一曲《榆盘洋芋花儿开》,神仙的生活也不过如此。

突出表现为生态系统产出下降,特别是渔业资源显著下降,长江干流的年捕捞产量由20世纪50年代的40多万t下降到现今的不足10万t,且鱼类资源小型化趋势明显。水生态系统结构受损,环境承载力和对污染的处理能力下降,局部江段和湖泊污染严重、水华时有发生,影响饮用水安全

In summary, both the new hemostatic compression device and the TR Band can efficiently achieve hemostasis following transradial coronary catheterization. However, fewer patients felt discomfort with application of the new hemostatic compression device. Pulse loss in the artery distal to the compression device, obstruction of venous reflux,and RAO were significantly lower with the new device.

ACKNOWLEDGEMENTS

We are grateful to the patients who participated in the study, and to the physicians and nurses who assisted.

REFERENCES

1. Johnson LW, Lozner EC, Johnson S, et al. Coronary arteriography 1984-1987: A report of the registry of the society for cardiac angiography and interventions. I. Results and complications. Cathet Cardiovasc Diagn 1989; 17:5-10.

2. Johnson LW, Krone R. Cardiac catheterization 1991: A report of the registry of the society for cardiac angiography and interventions (SCA&I). Cathet Cardiovasc Diagn 1993;28:219-220.

3. Hessel SJ, Adams DF, Abrams HL. Complications of angiography. Radiology 1981; 138:273-281.

4. Campeau L. Percutaneous radial artery approach for coronary angiography. Cathet Cardiovasc Diagn 1989; 16:3-7.

5. Kiemeneij F, Jan Laarman G. Percutaneous transradial artery approach for coronary stent implantation. Cathet Cardiovasc Diagn 1993; 30:173-178.

6. Kiemeneij F, Laarman GJ. Percutaneous transradial artery approach for coronary Palmaz-Schatz stent implantation. Am Heart J 1994; 128:167-174.

7. Kiemeneij F, Laarman GJ, Odekerken D, et al. A randomized comparison of percutaneous transluminal coronary angioplasty by the radial, brachial and femoral approaches: The ACCESS study. J Am Coll Cardiol 1997; 29:1269-1275.

8. Agostoni P, Biondi-Zoccai GG, de Benedictis ML, et al. Radial versus femoral approach for percutaneous coronary diagnostic and interventional procedures; Systematic overview and meta-analysis of randomized trials. J Am Coll Cardiol 2004;44:349-356.

9. Roberts JS, Niu J, Pastor-Cervantes JA. Comparison of hemostasis times with a kaolin-based hemostatic pad (QuikClot Radial) vs mechanical compression (TR Band) following transradial access: A pilot prospective study. J Invasive Cardiol 2017; 29:328-334.

10. Rathore S, Stables RH, Pauriah M, et al. A randomized comparison of TR band and radistop hemostatic compression devices after transradial coronary intervention. Catheter Cardiovasc Interv 2010; 76:660-667.

11. Pancholy SB. Impact of two different hemostatic devices on radial artery outcomes after transradial catheterization. J Invasive Cardiol 2009; 21:101-104.

12. Cubero JM, Lombardo J, Pedrosa C, et al. Radial compression guided by mean artery pressure versus standard compression with a pneumatic device (RACOMAP). Catheter Cardiovasc Interv 2009; 73:467-472.

13. Dai N, Xu D-C, Hou L, et al. A comparison of 2 devices for radial artery hemostasis after transradial coronary intervention.J Cardiovasc Nurs 2015; 30:192-196.

14. Fernandez RS, Lee A. Effects of methods used to achieve hemostasis on radial artery occlusion following percutaneous coronary procedures. JBI Database System Rev Implement Rep 2017; 15:738-764.

15. Nagai S, Abe S, Sato T, et al. Ultrasonic assessment of vascular complications in coronary angiography and angioplasty after transradial approach. Am J Cardiol 1999;83:180-186.

16. Pancholy SB, Bertrand OF, Patel T. Comparison of a priori versus provisional heparin therapy on radial artery occlusion after transradial coronary angiography and patent hemostasis(from the PHARAOH study). Am J Cardiol 2012;110:173-176.

Yong Wang,Jing Tang,Jingwei Ni,Xin Chen,Ruiyan Zhang
《Journal of Interventional Medicine》2018年第4期文献